N9030B PXA Signal Analyzer Service Guide 131
Instrument Messages
Condition Messages
159
13
Settings Alert; Analog Out
settings conflict
W
The user has manually set the Analog
Output under the Input/Output menu to
a setting that conflicts with the current
measurement. There will be no output
on the Analog Out port until this conflict
is resolved. In most cases, simply set
Analog Out to Auto for the optimal
setting.
159
13
Settings Alert;I/Q
mismatch:<Differential|Input
Z|Attenuation>
W
The impedance, differential, or
attenuation settings for the I and Q
channels do not match. For valid I+jQ
measurements the impedance and
differential settings should be the same
on both channels and the attenuation
should match within 1 dB
159
13
Settings Alert;Parm/data
mismatch
W
For
Bluetooth
, the detected
parameters did not match the data
161
14
Setting Modified
E
161
14
Setting Modified; Filter not
applied
E
The filter you have selected is larger
than the sampling frequency. You
should select a different filter.
Err#
Bit in
status
register
Message
Error or
Warning
More Information
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...