N9030B PXA Signal Analyzer Service Guide 503
Assembly Replacement Procedures
Reference Assembly
4.
. Remove cables W3, W18, and W31 from the
reference assembly
(1)
using the 5/16-inch wrench.
Figure 16-70
Reference Board Cables
5.
Remove the two screws
(2)
) attaching the Reference board to
the chassis.
Table 16-16
Item
Keysight Part Number
W3
W18
W31