232 N9030B PXA Signal Analyzer Service Guide
Analog/Digital IF Troubleshooting
A2 Analog IF Troubleshooting
11.
The analyzer should read 322.5 MHz at
−
23 dBm
±
Figure 7-2
A15 322.5 MHz Output
Reconnect W25 to A15J900.
IMPORTANT
Measuring at this location is for convenience. The 322.5 MHz signal is an
output at A15J900. Since the W25 cable connects from this output to the
input of the Analog I.F. assembly at A2J100, the small coaxial cable has not
been tested at the A2J100 end.
If the 322.5 MHz signal is not measuring the correct power level, refer to
Chapter 4, “RF Section Troubleshooting (RF/Microwave Analyzers)”
service guide.