N9030B PXA Signal Analyzer Service Guide 591
Functional Tests
Frequency Response (Flatness)
4.
Press
Mode
,
Spectrum Analyzer
,
Mode Preset
on the signal analyzer, and
press
Preset
on the synthesized sweeper.
5.
Set up the synthesized sweeper by pressing:
CW,
50,
MHz
Power level
, –4,
dBm
6.
Set up the signal analyzer by pressing:
Input/Output
,
More
,
Freq Ref In
,
External
FREQ Channel, Center Freq
, 50,
MHz
SPAN X Scale, Span
, 50,
kHz
AMPTD
Y Scale, Ref Level, 0,
dBm
7.
Adjust the synthesized sweeper output power for a power meter reading
of –10 dBm ±0.1 dB.
8.
On the signal analyzer, press
Single
.
9.
Press the
Peak Search
key on the signal analyzer to position the marker on
the peak of the signal.
10.
Refer to
Table 18-7, “Frequency Response (Flatness) Results.”
Enter the
amplitude of the signal displayed on the signal analyzer into the Meas
Amp
column of
.
11.
Enter the power meter reading into the Power
Meter
column of
12.
Tune the synthesized sweeper and signal analyzer to the next frequency
listed in
.
13.
Enter the power sensor calibration factor into the power meter.
14.
For frequencies 3.6 GHz and above, press
AMPTD Y Scale
, then
Presel
Center
to center the preselector filter for an optimum amplitude
measurement.
15.
through
and complete the remainder of
for the frequency range of your analyzer.
16.
Compute the measurement error (Meas
Error
= Meas
Amp
- Power
Meter
).
17.
Compute the flatness error normalized to 50 MHz:
(Meas
Error
- Meas
Error
@ 50 MHz)
18.
Enter the computed flatness error value into the Flat
Norm
column of
19.
Compare the value of Flat
Norm
to the test limit.
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...