1.
. Place the switch onto the bracket and replace the
five screws
(3)
). Torque to 9 inch-pounds.
2.
Place the switch/bracket into place into the chassis and replace the two
screws (
). Torque to 9 inch-pounds starting with the screw
closest to the front of the instrument.
3.
. Replace the semi-rigid cables W6, W8, and W56.
Torque to 10 inch-pounds.
4.
Replace the ribbon cable W7. Ensure locking tabs on the sides of the
connector are engaged.
5.
. Position the RF bracket onto the chassis and
replace the sixteen screws (
). Torque to 9 inch-pounds.
6.
Replace the front panel. Refer to the
replacement
procedure.
7.
Replace the instrument outer case. Refer to the
replacement procedure.