N9030B PXA Signal Analyzer Service Guide 521
Assembly Replacement Procedures
AIF/DIF Assembly
AIF/DIF Assembly
Removal
1.
Remove the instrument outer case. Refer to the
removal procedure.
2.
Remove the rear panel. Refer to the
removal procedure.
3.
Refer to
. Remove cables W26, W39, W42, and W47 from the
bottom of the DIF assembly.
4.
. Pull the AIF/DIF assembly part way out of the
chassis. Remove cables W25, W36, and W42 from the AIF assembly.
5.
The AIF/DIF assembly can be removed from the chassis by pulling straight
out the back.
Figure 16-84
DIF Cables