600 N9030B PXA Signal Analyzer Service Guide
Functional Tests
Scale Fidelity
8.
On the analyzer, activate the Marker Delta function by pressing
Peak
Search
,
Marker Delta
.
9.
Perform the following steps for each attenuator setting listed in the table
below:
a.
Select the next External attenuator setting.
b.
Press the
Restart
key to trigger a 10 sweep average.
c.
Enter the delta marker value into
d.
Check delta marker reading against the test limits.