478 N9030B PXA Signal Analyzer Service Guide
Assembly Replacement Procedures
RF Area (Options 544, 550)
Low Band Switch
Removal
1.
. Remove the ribbon cable W7.
2.
Remove the semi-rigid cables W6, W8, and W56 using the 5/16 inch
wrench. W6 and W56 are only present on analyzers without Option MPB
and/or LNP. If either of these options are present, the equivalent cables
would be W52 (for W6) and W54 (for W56).
3.
. Remove the two screws
(2)
(
) using the
T-10 driver. The low band switch
(1)
can now be removed from the
chassis.
4.
Refer to
. To separate the switch from the bracket, remove the
five screws
(3)
) using the T-10 driver.
Figure 16-45
Low Band Switch Cable Removal
If the analyzer has hardware for Options MPB and LNP installed, perform the removal procedure
first.