510 N9030B PXA Signal Analyzer Service Guide
Assembly Replacement Procedures
Rear Panel
Rear Panel
Removal
1.
Remove the instrument outer case. Refer to the
removal procedure.
2.
. Remove W37
(2)
cable from the rear panel using
the 9/16- inch nut driver. Remove W27
(3)
cable using the 5/16- inch
wrench.
3.
Using the T-10 driver, remove the sixteen screws
(4)
(
)
attaching the rear panel
(1)
to the chassis. The rear panel can now be
removed.
Figure 16-77
Rear Panel Removal
Replacement
1.
Reattach the W37 cable to the rear panel. Torque to 21 inch-pounds using
the 9/16-inch nut driver.
2.
Reattach the W27 cable to the rear panel. Torque to 10 inch-pounds using
the 5/16-inch wrench.
3.
Place the rear panel into position in the chassis. Replace the sixteen
screws (
) to attach the rear panel to the chassis. Torque to
9 inch-pounds.
4.
Replace the instrument outer case. Refer to the
replacement procedure.