558 N9030B PXA Signal Analyzer Service Guide
Post-Repair Procedures
Post-Repair Procedures
Figure 17-1
Alignment Data Wizard
5.
Follow the on-screen instruction to backup the calibration data to the
USB FLASH drive.
If the Instrument Software Application Will Not Run
1.
Connect a USB mouse to one of the instrument front panel USB ports.
2.
Connect a USB FLASH drive to the other front panel USB port.
3.
Using the mouse double-click on the Windows Explorer link on the
desktop and navigate to the following folder:
C:\Program Files\Agilent\SignalAnalysis\Physics
4.
Double-click on BackupAndRestore.exe and the window shown in
will appear.