N9030B PXA Signal Analyzer Service Guide 475
Assembly Replacement Procedures
RF Area (Options 544, 550)
Attenuators
Removal
1.
. To remove Attenuator A
(1)
or Attenuator B
(2)
,
remove the semi-rigid cables W2, W3, W4, or W6 attached to the
attenuator using the 5/16 inch wrench.
2.
Remove the ribbon cable attached to the attenuator.
3.
Remove the two screws
(3)
or
(4)
(
) for each attenuator
bracket to remove from the chassis using the T-10 driver.
4.
Remove the attenuator and magnetic shield
(5)
or
(6)
from the bracket
by removing the two screws (
).
Figure 16-44
Attenuators Removal
Table 16-11
Item
Keysight Part Number
W2
W3
W4
W6