124 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Condition Messages
Condition Errors 36 to 64, Calibration Needed or Failed
This series of errors corresponds to the bits in the
STATus:QUEStionable:CALibration register. The second column in the table
shows the corresponding bit.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These error numbers can be seen in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
For example, error 42 indicates that the “RF Alignment Failure” condition has
been detected, and error 1042 indicates that failure has been cleared.
Several bits in this register are “summary bits” for registers at a lower level.
There are no error messages associated with these bits; they exist only as
status bits, read with a STATus:QUEStionable:CALibration? event query or a
STATus:QUEStionable:CALibration:CONDition? query.
Note that these summary bits summarize the state and history of the event
registers at the lower level. This is true even for bits in the
STATus:QUEStionable:CALibration condition register. This means that:
The summary bits read by the STATus:QUEStionable:CALibration:CONDition?
query are true if any event bits are set in any of the :CALibration sub-registers
:SKIPped, :EXTended:NEEDed or :EXTended:FAILure.
The summary bits read by the STATus:QUEStionable:CALibration? event query
are true if any event bit has undergone a false-to-true transition with the
PTRansition filter set, or a true-to-false transition with the NTRansition filter
set, in any of the :CALibration sub-registers :SKIPped, :EXTended:NEEDed or
:EXTended:FAILure.
18
6
unused
20
7
unused
22
8
unused
24
9
unused
26
10
unused
28
11
unused
30
12
unused
32
13
unused
34
14
unused
Err#
Bit in
status
register
Message
Error or
Warning
More Information
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...