Operating the TSG-3800
Default Parameter Values
Provisioned SQL
STU
ST1/STU/ST2/ST3
SSM Mode
AUTO
PROV/AUTO
SSM Alarm Mode
ALARM
Ignore/Report/Alarm/Fail
Input Priority
1.1=1 / 2.1=2 /
all others=0
0-4
E1 SSM Bit
4
4/5/6/7/8
Note:
Input Module is either T1 or E1 but either also has option for CLOCK input signals.
INPUT ERROR THRESHOLDS
AIS Clear/Error
(2/10)
AIS Mode
Fail
Ignore/Report/Alarm/Fail
BPV Clear/Error
(2/10)
BPV Mode
Alarm
Ignore/Report/Alarm/Fail
LOS Clear/Error
(2/10)
LOS Mode
Fail
Ignore/Report/Alarm/Fail
OOF Clear/Error
(2/10)
OOF Mode
Alarm
Ignore/Report/Alarm/Fail
CRC Clear/Error
(2/10)
CRC Mode
Alarm
Ignore/Report/Alarm/Fail
FREQUENCY THRESHOLDS
100s Limit 1
8000ns - Fail
1000s L1
800ns - Fail
10000s L1
80ns - Fail
100s Limit 2
4000ns - Alarm
1000s L2
400ns - Alarm
10000s L2
40ns - Alarm
MTIE THRESHOLDS
100s Limit 1
432 ps/s - Fail
1000s L1
864 ps/s - Fail
10000s L1
1728 ps/s - Fail
Table 3-6. Default Software Configuration (Continued)
Setting
Factory Default
Range
User Settings