96
System Description
How the Echelle Spectrometer and SCD Work
Introduction
The Avio 500 optical design combines an echelle polychromator with a solid-state
detector in an integrated system optimized for ICP-OES. The echelle
polychromator produces a compact two-dimensional spectrum that is focused onto
the solid-state detector. The Segmented-array Charged-coupled device Detector
(SCD) simultaneously measures emission lines for the analytes and the spectral
background. On the silicon surface of the detector is a collection of linear detector
segments strategically located to measure the most important ICP emission lines.
Random access to data for individual emission lines ensures rapid analysis.
The Avio 500 is a dual detector system. The UV detector covers an extended
ultraviolet wavelength range from 165 to 403 nm. The VIS detector covers the
visible wavelength range from 404 to 782 nm.
The optical system consists of the transfer optics, input module, and UV module
and a VIS module.
Transfer Optics
The torch emission is directed to the entrance slit by the transfer optics. A
computer-controlled toroidal mirror can be rotated both vertically and horizontally.
This computer-controlled movement allows the user to adjust the viewing height
and the viewing position. For more information on the transfer optics, see the
diagrams for radial and axial view Figure 3-2 see page 80 and Figure 3-3 page 81.
An automatic shutter mechanism (not shown in optical diagrams) minimizes the
UV exposure on the first toroidal mirror and the rest of the optical system thus
extending its useful lifetime. The shutter mechanism also includes a mercury pen
lamp. This pen lamp can be used to check the wavelength calibration of the system.
Input Module
The input module consists of three components: the computer-controlled slit
mechanism, the parabolic collimator mirror and the echelle grating. The slit
mechanism contains three slit openings: the “High Resolution” slit, the “Normal
Resolution” slit, and the “Low” (High Energy) slit. After the light from the plasma
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......