67
Location
Location
Typically, the Avio 500 is positioned with the computer and printer on the bench
next to the instrument or on an adjacent table. The computer and printer must not
be placed on the instrument. A bench for the instrument is available (Part No.
N0782060).
The instrument must be placed under the exhaust vent for the torch compartment
and near the electrical, water, and gas supply points.
Cooling Water Requirements
A water supply is required to dissipate heat from the oscillator.
Recirculating System
A recirculating system (chiller) must be used. The requirements for the chiller are:
•
Cooling Capacity at 20 °C is 2000 watts
Temperature Stability ±1
o
C
Pump Rate 1 US gal/min. at 55 psi max (45 psi min).
Coolant: coolant fluid with corrosion inhibitor
Caution
Attention
If a longer hose distance is required for the chiller location, it may be necessary
to increase the hose inner diameter to accommodate the specification of
no
greater than 15 psig
at the instrument water outlet. The hose inner diameter
should also be sized large enough to accommodate the specification of 45 to
80 psig at the water inlet. The hose also needs to be insulated sufficiently to
maintain a temperature at the instrument water inlet between 15 °C and 25 °C.
Si une distance de tuyau plus long est nécessaire pour la localisation de
refroidissement, il peut être nécessaire d'augmenter le diamètre intérieur du
tuyau pour recevoir le cahier des charges
ne dépassant pas 15 psig
à la sortie
d'eau de l'appareil. Le diamètre intérieur du tuyau doit également être
dimensionnée assez grande pour recevoir la spécification de 45 à 80 psig à
l'entrée d'eau. Le tuyau doit également être suffisamment isolé pour maintenir
une température à l'entrée d'eau de l'instrument entre 15 ° C et 25 ° C.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......