Troubleshooting
Performance Problems: Troubleshooting Table
Table 6-16.
Performance Problems: Troubleshooting Table
Problem
Possible Cause(s)
Corrective Action
RSD
(Relative
Standard
Deviation)
greater than
1%.
The sample introduction
system may not be set up
correctly or may require
maintenance.
.
Does the sample have a high
viscosity or high percentage
of dissolved solids?
Try increasing the integration time. Refer to
the Syngistix Software Guide on the
Syngistix Document Pack CD for details
.
For high dissolved solids, try the GemCone
spray nebulizer.
Add 0.05% surfactant to your samples,
standards, and blanks.
Try running other samples with normal
characteristics to determine if the sample
matrix is really the problem.
The standard and blank may need to be
matrix-matched. Or, internal standardization
may be required.
Look for carryover – check if
concentrations of replicates
increase from one to the next.
Increase the Read Delay to flush the sample
introduction system completely. Increase the
rinse time if using an autosampler. Prepare
the rinse solution in the same acid matrix
used for the standards, samples, and blank.
Refer to the Syngistix Software Guide on the
Syngistix Document Pack CD for details
.
RF power and/or nebulizer
argon flow may require
adjustment.
Adjust as necessary. Also, check that you are
using the recommended settings for your
instrument.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......