Installation
Automatically Aligning the Torch Viewing Position
This procedure adjusts the torch viewing position for the highest signal intensity.
Perform this procedure when:
•
the instrument or software is first installed
•
the instrument is moved to a new location
•
the torch is removed or replaced
•
the plasma induction plates are replaced
1. Switch on the computer and spectrometer, and if required, the autosampler.
2. Switch on the exhaust ventilation system.
3. Start the ICP Syngistix software.
4. In the
Options
menu, click on
Autosampler
, and make sure that the software
is configured for the autosamper tray you are using.
5. In the
Tools
menu, click on
Plasma Control
and ignite the plasma. Allow the
torch to warm up for half an hour before you start the alignment.
6. Prepare a solution containing 1 ppm manganese.
7. In the
Tools
menu, click on
Spectrometer Control
.
The Spectrometer Control window appears.
8. Select
Axial
or
Radial
for the 8X00 DV. We recommend that you first do the
alignment in one viewing mode, then select the other viewing mode and repeat
the procedure.
9. To view spectra collected during the procedure, open the
Spectra Display
window.
10. Click on
Align View
in the Spectrometer Control window to open the
Align
View
dialog.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......