68
Preparing Your Laboratory
A PolyScience Recirculating Chiller meets these requirements and is
recommended for the instrument. The PolyScience
®
WhisperCool
TM
chiller is
available through PerkinElmer in the following two configurations:
•
230V, 60 Hz (Part No. N0772046)
•
240V, 50 Hz (Part No. N0772045)
For detailed information on PolyScience
®
WhisperCool
TM
chiller electrical
requirements see the following section,
Electrical Requirements
.
The PolyScience
®
WhisperCool
TM
chiller comes with its own Instruction Manuals.
The headquarters for PolyScience Instruments is located at:
PolyScience Instruments, Inc.
6600 West Touhy Ave
Niles, Il 60714
USA: (800) 229-7569
Worldwide: (847) 647-0611
Fax: (847) 647-1155
Electrical Requirements
Avio 500
•
The Avio 500 ICP-OES is equipped with an IEC 309 250 V 16/20 A 2-pole
plus protective earth plug (Hubbell® Part No. C320P6SVL, PerkinElmer Part
No. 09997530) that inserts into a receptacle (Walther Part No. 410306,
PerkinElmer Part No. 09290304). Alternative surface mount receptacle
(Walther Part No. 111306, PerkinElmer Part No. 09290305). Both receptacles
are shipped with the instrument.
Grounding circuit continuity is vital for safe operation of this instrument.
Grounding is accomplished by use of an IEC60309 instrument power plug
furnished by the factory and not to be removed during or after installation.
Avertissement: Mise à la terre la continuité du circuit est essentiel pour une
utilisation sûre de cet instrument. La terre est effectuée par l'utilisation d'une
prise d'alimentation de l'instrument IEC60309 fourni par l'usine et ne doit pas
être retiré pendant ou après l'installation.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......