175
Automatically Aligning the Torch Viewing Position
To perform UV channel wavelength calibration:
1. Aspirate the Wave Cal solution (Part No. N0582152).
2. Select
UV
, then click on
Wavelength Calibration
in the Spectrometer Control
window.
The UV calibration may take several minutes. You may view the spectra used
for calibration by clicking on
Spectra
in the toolbar. When the UV calibration
is complete, the UV coefficients and RMS (Root Means Square) value will be
displayed in the Wavelength/Calibration window. These values should meet
the following specifications:
If these specifications are not met, repeat the wavelength calibration and make sure
all peaks can be seen in the Spectra Display window. If these specifications are still
not met, a complete optical alignment may be required. Contact a PerkinElmer
Service Representative.
To perform VIS channel wavelength calibration:
1. Aspirate VIS Wavecal mix (Part No. N9302946).
2. Select
VIS
, then click on
Wavelength Calibration
in the Spectrometer
Control window.
The VIS calibration may take several minutes. You may view the spectra used for
calibration by clicking on
Spectra
in the toolbar. When the VIS calibration is
Note
The spectrometer will automatically change to radial viewing to prevent
saturation of the detector.
UV Coefficients and RMS
Specification
First Coefficient
absolute value < 2.0
Second Coefficient
absolute value < 2.0
Third Coefficient
absolute value < 8.0
RMS
(measures variability of
wavelength offsets)
< 2.0
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......