230
Maintenance
Replacing the Plasma Induction Plates
Plasma Induction Plates
Regularly inspect the plasma induction plates. They should not show any signs of
deforming or pitting.
If your lab has high humidity, check the plasma induction plates to make sure they
are dry. Use a soft dry cloth to dry the plasma induction plates if necessary. Contact
your PerkinElmer Service Representative to replace and adjust the plasma
induction plates.
Replacing the Plasma Induction Plates
The plasma induction plates may need to be replaced occasionally as a result of
overheating, pitting, or arcing. Contact your Perkin Elmer Service Representative
to replace the plasma induction plates.
Table 5-3.
Replacement Parts: Torch Module
Part No.
Description
09992731
Axial Window
N0810346
Radial Purge Window
N0810128
Torch Body Assembly
N0810606
Quick Change Torch Module-Scott Spray Chamber
N0810607
Quick Change Torch Module-Cyclonic Spray Chamber
N0791180
Alumina Injector 0.80mm
N0791181
Alumina Injector 1.2 mm
N0791182
Alumina Injector Straight 1.2 mm
N0791183
Alumina Injector, 2.0 mm ID, 150 mm LG (standard injector)
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......