233
Windows
Removing and Replacing the Windows
Normally you should remove the axial window before you remove the radial
window. This releases the pressure in the spectrometer and helps avoid objects or
contamination falling into the radial snout.
Removing and Returning the Axial Purge Window
Tool Needed:
•
Purge window aluminum
ring removal tool (Part No. N0790448)
1. Turn off the plasma.
Caution
Attention
These windows are important parts of the optical system. Handle them
carefully, as you would any sensitive optical component. Do
not
directly
touch the windows. Always remove the axial window before the radial
window and refit the radial window before the axial window. This will
help avoid objects or contamination falling into the radial snout.
Ces fenêtres sont des éléments importants du système optique. Manipulez-
les avec soin, comme vous le feriez pour tout composant optique sensible.
Ne touchez pas directement les fenêtres. Retirez toujours la fenêtre axiale
avant que la fenêtre radiale et remonter la fenêtre radiale devant la fenêtre
axiale. Cela permettra d'éviter des objets ou la contamination de tomber
dans le museau radial.
Caution
Attention
Wear
cotton gloves
when you handle the torch glassware
.
Porter des gants de coton lorsque vous manipulez la verrerie de la torche.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......