240
Maintenance
5. With the tool in place rotate the tool to loosen the stainless steel ring.
6. With the ring now loosened carefully unscrew the radial purge window mount
the rest of the way by hand.
7. Remove the ring that holds the radial purge window in the mount
8. Lift the radial window and ring out of the mount — take care if the
spectrometer purge is set to high and you have not removed the radial window,
the pressure inside the spectrometer may force out the radial window.
9. If the window is dirty clean it by following the procedure Cleaning the
10. If the window is damaged, replace with a new radial purge window (Part No.
N0810346).
11. Return the new radial purge window in the stainless steel ring to the
instrument.
12. Carefully screw it into place.
13. Use the tool to tighten down the stainless steel ring and radial purge window
into place.
14. Return the Shear Gas Nozzle. See Returning the Shear Gas Nozzle on page
15. Turn on the plasma.
Cleaning the Windows
Note
Do
not
rub the surfaces of the windows before you have cleaned them as
described below. The more you rub the surface of the windows, the more
chance there is of scratching them. You must discard scratched windows.
Allow to dry completely before reinstalling. If there is any moisture on the
inside of the window, it can degrade the UV performance and/or damage
the optics.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......