122
Installation
The PolyScience is available through PerkinElmer in the following two
configurations:
208/230V, 60 Hz
220/240V, 50 Hz
Connecting the Chiller to the Instrument
Connect the instrument to the chiller with hoses. The direction of the flow through
the system can be controlled by the way the hoses are connected to the chiller. The
"INLET" port will draw liquid into the chiller; the "OUTLET" port will pump
liquid out. See the following figure for the location of the cables.
1. Locate the instrument Return/Output coolant line (Part No. N0770342) and
connect to the chiller "INLET" fitting. The instrument water outputs have a
fitting and tubing to connect them to the return or drain. Use 3/8-in. space
collars (Part No. 09920584) at the instrument. The following figure shows the
water supply connections at the instrument.
2. Locate the instrument Input coolant line (Part No. N0770341, uses 3/8-in.
Swagelok fittings) and connect to the chiller "OUTLET" fitting. Use 3/8-in.
space collars (Part No. 09920584) at the instrument. The following figure
shows the chiller connections. The following figure shows the water supply
connections at the instrument.
3. Make sure that the hoses and fittings are tight and that there are no bends or
crimps in the hoses.
Note
Teflon tape (PTFE tape) should be used for all pipe-thread
(NPT) fittings.
Use of Teflon tape can prevent leaks.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......