Troubleshooting
Are there deposits inside the end cap or spray chamber?
Clean out deposits. See Chapter 5, Maintenance for recommended cleaning
procedures (
Table 5-1 on page 204
).
Torch Assembly
Are the torch fittings for the gas inlets secure?
Finger-tighten the fittings. Do not overtighten them.
Is the quartz torch cloudy or dirty?
Check for deposits, particularly if running organics, alloys, or samples with high
dissolved solids. See
Cleaning the Torch on page 217
Is the injector dirty? Is it deformed due to overheating?
The injector may be clogged and require cleaning. If you are running organics,
check for carbon buildup. If running samples with high dissolved solids, check for
deposits. See
Cleaning the Torch on page 217
If the torch requires disassembly, check the O-rings. Are they cracked or worn? Is
the injector adapter discolored?
Replace worn parts. See
Replacing the Torch on page 218
Purge Window
Is the purge window cloudy or dirty?
Rinse with deionized water and dry with a soft cloth. Or, replace if necessary. Do
not touch the window. If oils from the skin are deposited on the window, this can
cause inaccurate results. A 10% solution of hydrofluoric acid may be used to clean
the window. Observe the following warning statement. Soak the window briefly,
then rinse thoroughly.
Has UV performance dropped?
Even if the purge windows look clean a drop in UV performance indicates that the
windows should be cleaned or replaced. First rinse with deionized water and dry
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......