293
Peristaltic Pump
5. Tighten the adjustment screw slowly (turn clockwise) until you just see a
smooth flow of bubbles. Add an additional three more turns.
6. Turn off the pump. The tension is now correctly set. No bubbles should be seen
in the sample tubing once the sample tube is kept in the solution.
Cleaning the Pump
Periodically inspect the peristaltic pump rollers to make sure they are clean and
move freely. Clean the exterior of the pump with a cloth moistened with water.
The pump head on the peristaltic pump can be removed and cleaned if necessary.
A replacement pump head is available (Part No. 09220266).
Removing the Pump Head on the Peristaltic Pump
To remove the pump head on the peristaltic pump:
1. Remove the pump tubing.
2. Unscrew the large screw in the center of the pump head. Place the screw and
washer in a secure place.
3. Lift the pump head off. To clean the pump head, use water or a mild solvent
and dry thoroughly.
4. Return the pump head to the pump. Orient the pump head by aligning the notch
on the bottom of the pump head with the pin on the shaft. The pump head must
click in this slot for a proper fit. Replace the washer and tighten the screw.
Caution
Attention
If a spill occurs on the pump, immediately clean it up.
A corrosive spill
could damage the pump.
Si un déversement se produit sur la pompe, nettoyer immédiatement
vers le haut.
Un déversement corrosif pourrait endommager la pompe.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......