165
Fitting and Connecting the Autosampler Sampling Probe
Fitting the Sampling Probe
1. If not already done, carefully insert the sampling probe guide from above into
the sampling probe holder on the autosampler arm and push it down fully, as
shown in the following figure.
2. Carefully insert the sampling probe from above into the probe guide.
3. Push the sampling probe down until about 20 mm protrudes from the bottom
end of the probe guide, then position the O-ring on the probe to set this
position, as shown in the following figure.
This is not the final position for the sampling probe. The probe can be adjusted
when the rinsing port has been set up. See the following procedure,
the Sampling Probe Height on page 167
Note
The standard sampling probe is a PTFE-lined epoxy polymer tube. This
polymer is resistant to the acids, alkalis and organic solvents normally used
in atomic spectroscopy. If your samples contain an organic solvent that
attacks this polymer, a PTFE-lined stainless steel sampling probe is
offered as an option (Part No. B3000152).
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......