117
Moving the Instrument
Moving the Instrument
If you must move the instrument any great distance, especially if the instrument
may be subject to vibration or jolts
contact a PerkinElmer Service
Representative to assist you in moving your system
.
If you are only moving the instrument within the lab or to a laboratory nearby you
may use the following procedure.
•
Remove the torch module.
•
Remove the chiller.
•
Remove any accessories, for example an autosampler.
•
Disconnect all gas and water lines.
•
Remove the computer and printer.
•
Attach the handles.
•
Lift the instrument onto a movable table or put it in the wooden platform
originally shipped with the instrument. See the Unpacking Instructions (Part
No. 09931199).
•
Lift into position at the new location. Make sure that the new location complies
with the laboratory requirements; see
Preparing Your Laboratory on page 59
•
At the new location reconnect the torch module, the chiller, any accessories,
gas and water lines, the computer and the printer; see the procedures later in
this chapter.
Caution
Attention
I
f moving the Avio 500 will subject the instrument to any freezing temperatures you
must
contact a PerkinElmer service engineer to assist you in the move. The
PerkinElmer service engineer will flush all traces of cooling water from the RF
generator to prevent freezing of RF generator components.
Si le déplacement du Avio 500 assujettira l'instrument à des températures de
congélation, vous devez contacter un technicien de maintenance PerkinElmer pour
vous aider dans le mouvement. L'ingénieur de service PerkinElmer rincera toutes
les traces de l'eau de refroidissement du générateur RF pour empêcher le gel des
composants RF du générateur.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......