Maintenance
Mira Mist Nebulizer
This device operates on compressed gases.
Appropriate care
must be taken. If in doubt about correct operating procedures, call
an experienced operator. Burgener Mira Mist Nebulizers require
45 - 55 psi to have a 1 liter per minute of Argon gas flow, so the
operating pressures are in the range of 25 - 45 psi, depending on
the torch optimum flow rate.
Mira Mist nebulizers will pulse if the pump cannot deliver
constant sample flow: Change your pump tubing often, or use a
surgeless pump if possible.
Avertissement: Cet appareil fonctionne sur les gaz
comprimés.
Des soins appropriés doivent être prises. En cas de
doute sur les procédures de fonctionnement, appeler un opérateur
expérimenté. Burgener Mira Mist nébuliseurs nécessitent 45 - 55
psi d'avoir un 1 litre par minute d'écoulement du gaz argon, de
sorte que les pressions de fonctionnement sont de l'ordre de 25 -
45 psi, en fonction du débit optimal de la torche.
Mira Mist nébuliseurs flashe si la pompe ne peut pas fournir un
débit d'échantillon constant: Changez votre tuyau de pompe
souvent, ou utiliser une pompe surgeless si possible.
!
Warning
!
Warning
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......