34
Safety Practices
Handling of the Instrument
Contact a PerkinElmer service engineer for assistance in installing or
evaluating the system after moving.
The Avio 500 weighs 163 kg (359 pounds),
or 203 kg (447 pounds) with the shipping container. If the instrument needs to be
moved please contact PerkinElmer Service.
Laboratory Ventilation
Toxic combustion products, metal vapor, and ozone can be generated by the
system, depending upon the type of analyses. Combustion products vary with the
type of sample analyzed. If the sample compartment door is opened, a maximum
of 0.9 liters of Argon may escape during the first minute. After that, 0.1 liters/
minute of Argon will continue to escape while the door is open.
An exhaust venting system is always required to remove those gases which are
generated during the normal operation of the instrument.
An efficient ventilation system must be provided for the instrument.
Instrument Safety Practices
This section describes potential hazards and recommended safety practices. You
should thoroughly review this information.
This equipment requires no specified inspection or preventive maintenance
to ensure the continuous functioning of its safety features.
Position the Avio 500 so that is can be easily unplugged.
Caution
Attention
The protection provided by this equipment may be impaired if
the equipment is used in a manner not specified by PerkinElmer.
La protection fournie par cet équipement risque d'être moins efficace
si l'équipement fait l'objet d'une utilisation différente de celle
mentionnée par PerkinElmer.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......