82
System Description
A flow of shear gas (compressed air or nitrogen) is directed at the plasma to "shear"
off the tip of the plasma discharge. The shear gas minimizes the effects of self-
absorption by forcing cooler atoms from the tip of the plasma out of the optical
path. The shear gas also keeps the purge window cool. The shear gas flow is turned
on automatically when the plasma is ignited.
Major Components
The Avio 500 consists of two major components: the spectrometer and the ICP
Source. The Avio 500 consists of two detectors. The plasma source is a dual-view
(can be viewed either axially or radially) RF generator. The sample introduction
system includes a built-in peristaltic pump and a "quick change" removable torch.
The removable torch includes the nebulizer and spray chamber. The nebulizer and
spray chamber are located outside of the instrument for easy accessibility and to
avoid being heated by the RF generator. The entire sample introduction system is
adjustable from outside the torch box while the RF generator is running. The torch
pneumatics includes a mass flow based nebulizer gas, a closed-loop pressure
restrictor based AUX and plasma gases.
Features of the Avio 500 include:
•
improved analytical performance especially in the low UV spectra
•
simultaneous background correction
•
solid-state generator for improved reliability
•
smaller bench top instrument
The Spectrometer and the ICP Source is further divided into the assemblies
described below.
The Spectrometer consists of the following assemblies:
•
Optical Tub assembly
(a) the Transfer Optics module
(b) the Input module
(c) the UV Detector module
(d) the VIS Detector module
•
Spectrometer Electronics assembly
•
Spectrometer Pneumatics assembly
The ICP Source consists of the following assemblies:
•
RF Generator and Torch Compartment assemblies
•
RF Control Electronics
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......