283
Mira Mist Nebulizer
Sample Capillary Tubing
The capillary tubing is attached with UpChurch® 10/32 chromatographic fittings.
We supply 0.043 OD X 0.017 ID Poly Capillary tubing (Part No. N0777111), but
you may use any other tubing that fits an UpChurch 10/32 fitting. We recommend
that you use a 0.017" ID or smaller ID capillary tube for the sample feed. This
should catch any particles before they get into the nebulizer. It is much safer &
easier to replace the capillary tubing than to clean the nebulizer.
The Mira Mist Nebulizer body is PEEK, and the inner capillaries are Teflon.
Generally it will not break. If it is dropped such that the tip is deformed, then it will
be irreparably damaged. If it continues to operate after being dropped, then it has
not been affected, and its use can be safely continued.
Gas Line
The gas line is attached with UpChurch® 10/32 chromatographic fittings (Part No.
N0777113). We supply 2 mm OD X 0.039 ID Teflon tubing. A gas line filter is
NOT
included in the nebulizer. Any particles from the gas line will destroy the
Caution
Attention
The gas orifice is at the very tip of the nebulizer. It is made of Teflon which
is
VERY SOFT
. This tip is very easily damaged and should
NEVER
be
touched with fingers, tissues, or anything else. If the tip is accidentally
touched and the nebulizer continues to operate, then it is still functional,
and its use can be safely continued.
L'orifice de gaz est à la pointe du nébuliseur. Il est fait de téflon qui est très
doux. Cette astuce est très facilement endommagé et ne doit
JAMAIS
être
touché avec les doigts, les tissus, ou toute autre chose. Si la pointe est
accidentellement touché et le nébuliseur continue à fonctionner, alors il est
encore fonctionnel, et son utilisation peut se poursuivre en toute sécurité.
Note
The PEEK Mira Mist body has 10/32 threads for the gas and capillary
fittings. Gently tighten until snug for the sample line. For the gas line, if
there seems to be any leakage, then tighten
HARD
. PEEK is very strong.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......