Autosampler Problems
Autosampler Problems
Table 6-19.
Autosampler Problems
Problem
Possible Cause(s)
Corrective Action
Autosampler does
not respond
There may be a
communication problem.
Use the
Instrument Diagnostics
window to reset the autosampler.
The software may not be
configured for the
autosampler you are using.
Select the
Instrument tab -> Devices
click on the
Autosampler
tab and
check the autosampler configuration.
If the wrong autosampler model is
shown exit, Syngistix and run the
Syngistix Reconfiguration
program
from the Windows start menu to
change the autosampler model.
One of the cables may have
been disconnected or the
autosampler may have been
turned off.
Check the power cable and the cable
connecting the autosampler to the
computer. Make sure the power is on.
The pins on the autosampler
cable connector may be
broken.
Check the connector. If the pins are
broken, order a replacement cable.
Autosampler does
not go to the correct
location.
The software may not be
configured for the correct
autosampler tray.
Autosampler tray is selected by going
to
Instrument
tab and selecting
Devices
and then select the
Autosampler
tab. In the
Autosampler Setting
field in the
Autosampler
tab select the tray type
from the browse window.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......