179
Setting Instrument Parameters
There is a three slot torch for organics (Part No. N0790249) that can be ordered.
This is the best torch to use for axial viewing for samples with organics or high
dissolved solids. Using the long torch will extend the torch life since the long torch
does not have a slot.
For more information on the above parameters, refer to the
Syngistic for ICP Guide
on the Syngistic for ICP CD.
Important Considerations for Dual Viewing
Optimum conditions will depend on the type of analyses being performed;
however, in general, the following considerations are important for axial viewing.
Settings are given for the Scott-type spray chamber and cross-flow nebulizer.
The following parameters are set in the method:
•
Auto integration times are recommended, using the following settings:
Auto Int Samp Time:
Min: 2 sec, Max: 10 sec
(Auto Integration Sampling Time)
•
Normal resolution is recommended.
•
Check that the read delay, wash time, and rinse times you are using are
sufficient. In general, either longer times are required or a fast pump speed. In
addition, check that the sample waste is being drained properly to avoid
solution filling the torch.
All integration parameters are set in the method editor.
Auto Integration Threshold:
2.0 sec recommended (default setting)
The Auto Integration Threshold sets a level for the integration. When the
software does a "snapshot" (called a pre-shot) for each solution, it will
determine the sampling time and adjust it based on the threshold. If the time is
greater than or equal to the threshold, the maximum sampling time will be
used. If it is less than the threshold, the minimum sampling time will be used.
In addition, see the following operating considerations:
Note
A Max time of 100 seconds may be required for the ultimate detection limit
work.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......