327
Performance Problems: Troubleshooting Table
Problem
Possible Cause(s)
Corrective Action
Undetectable
emission
signal
The sample introduction
system may not be set up
correctly, or may require
maintenance.
.
The shutter may be closed.
Check the shutter position set in the software.
Refer to the Syngistix Software Guide on the
Syngistix Document Pack CD for details
.
Memory
effects
Spray chamber may not be
being drained properly, or
may need cleaning.
Check drain fittings and drain tubing. Clean
the spray chamber if necessary.
Cannot get
suggested
detection
limits
The sample introduction
system may not be set up
correctly, or may require
maintenance.
.
The BEC may be high.
Run the BEC test. Refer to the Syngistix
Software Guide on the Syngistix Document
Pack CD for details
.
Look for carryover – check if
concentrations for the sample
replicates increase from one
to the next.
Increase the Read Delay to flush the sample
introduction system completely. Refer to the
Syngistix Software Guide on the Syngistix
Document Pack CD for details
.
The standard and blank may
need to be matrix-matched.
Prepare the standard and blank in the same
acid matrix used for the samples.
The integration time may be
too short.
Increase the minimum and maximum times
for integration. Refer to the Syngistix
Software Guide on the Syngistix Document
Pack CD for details
.
Table 6-16.
Performance Problems: Troubleshooting Table
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......