267
GemTip Cross-Flow Nebulizer
5. To reassemble the nebulizer make sure that the O-ring is completely seated on
the base of the argon tip, then insert the argon tip into the argon inlet.
6. Check the condition of the argon inlet tubing (Part No. 09985723). If the tubing
is worn, replace it as follows. See the next figure.
a. Obtain a new piece of 1/8-in. O.D. Teflon tubing (Part No. 09985723).
b. Using a razor blade, make a straight cut on one end of the tubing. Wire cutters
or scissors do not make acceptable cuts.
c. Insert a nut (Part No. 09920545) and ferrule (Part No. 09920515) on the newly
cut end of the tubing, with the ferrule oriented as shown in the figure below.
Figure 5-29
Installing new tubing for the argon inlet. Cross-Flow End Cap, Part
No. N0780546.
7. With the tubing and ferrule assembled as shown in the figure above, screw the
nut into the argon inlet making a finger-tight connection. Do
not
use any tools
to tighten the nut.
8. Examine the two GemTips in the end cap. These tips should almost butt up to
each other at an angle of 90
.
Item
Description
1
Ferrule (Part No. 09920515)
2
Nut (Part No. 09920545)
3
1/8 inch O.D. Tubing (Part No. 09985723)
3
2
1
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......