118
Installation
Connecting the Gases and Cooling Water
After the instrument has been moved into its position, it can be connected to the
various services in the laboratory.
Connecting the Gases
Use two wrenches when making Swagelok connections. Use a space collar (Part
No. 09920584, supplied in the hose kit) at each Swagelok connection. Tighten the
Swagelok nut until you can no longer turn the knurled space collar.
Only if you
cannot locate a space collar,
tighten the nut until it is finger-tight, then turn it an
additional 1/4 turn with a wrench. Pre-swaged fittings should be turned 1/8 turn
past finger-tight. Do not over-tighten fittings. Check for leaks with all the
connections made.
Connect the gases to the instrument as described in the following procedures.
Connecting the Nitrogen Supply
1. Connect the instrument to the nitrogen supply using the nitrogen hose (Part No.
N0690275, uses 1/4-in. Swagelok fitting) using a 1/4-in. space collar (Part No.
09920584) on both ends of the tubing. Refer to the following figure.
2. Set the nitrogen supply regulator between 275 kPa and 825 kPa (2.75 to 8.25
bar or 40-120 psig).
3. Check for leaks using a proprietary leak testing fluid for high purity gas lines.
Note
All gas hose connections use 1/4 inch Swagelok fittings.
Caution
Attention
The pneumatic tubing kinks easily.
Install it so that it is less likely to twist, fold and
kink
.
Les kinks de tubes pneumatiques facilement. Installez-le de sorte qu'il est moins
susceptible de se tordre, plier et kink.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......