262
Maintenance
1. If you have not already done so, remove the nebulizer/end cap assembly from
the spray chamber.
2. Disconnect the sample capillary tubing by removing it from the pump tube.
3. Disconnect the nebulizer argon tubing at the gas fitting. Do
not
disconnect
nebulizer argon tubing from the end cap because the GemTip could fall out.
4. Inspect the inside of the end cap. If metal or salt deposits are found, clean the
cap using a dilute acid solution.
Removing the Sample Tip from the End Cap, N0780546
1. Loosen the nut on the sample inlet
.
Remove the nut along with the tubing piece
and ferrule. See the next figure.
2. Using your fingers, push the sample tip out from the inside of the end cap,
being careful not to damage it. Do
not
use any metal tools to remove the
nebulizer tips because this may damage the tips or end cap.
3. Inspect the sample tip. Clean the tip using soap and water or a dilute solution
of nitric acid. An ultrasonic bath may be used.
If you need to replace the sample tip, be sure to replace it with a clear GemTip.
GemTips must be purchased in a set (Part No. N0780676). They are not
available separately.
4. Inspect the O-ring (Part No. 09921045), replace if it is worn or cracked.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......