66
Preparing Your Laboratory
Table 2-1.
Avio 500 ICP-OES Cables and Hoses Lengths
* Part of Tubing Assembly Kit (Part No. N0770183) TUBING ASSY – JAZ
EXTERNAL
Computer and Printer
Refer to the manuals supplied with your computer and printer for dimensions.
PolyScience Chiller
The PolyScience
®
WhisperCool
TM
(or equivalent) is usually located on the floor,
to the
right side of the instrument. Its dimensions are 36.5-cm (14.4 in.) wide, 67.3-
cm (26.5 in.) deep and 61.0-cm (24 in.) high. It weighs 80 kg (178 lbs.). The chiller
has air intakes and exhausts that need to be unobstructed. The power is 2200 watts.
It requires a minimum of 45.7 cm (18 in.) on all sides for adequate ventilation. The
coolant hoses restrict the distance that the chiller can be placed away from the
instrument (hose length = 3.6 meters or approximately 12 feet).
Part Number
Description
Length
N0810500
Avio 500 Power Cable Assembly
2.4 M (8 Ft)
09290332
Ethernet Cable
Instrument to Computer
4.57M (15 Ft)
09290798
USB Cable for Plasma Camera
Instrument to Computer
3M (9.84 Ft)
N0690274*
Argon Supply Hose
6M (20 Ft)
N0690275*
Nitrogen Supply Hose
6M (20 Ft)
N0770175*
Chiller Remote Cable
3.8M (12.4 Ft)
N0770341*
Water Supply Hose
Chiller to Instrument
2.7M (12 Ft)
N0770342*
Water Return Hose
Chiller to instrument
2.7M (12 Ft)
N0770348*
Shear Gas Supply Hose
2.7M (12 Ft)
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......