Peristaltic Pump
6. Insert the beveled end of the nebulizer capillary tubing into the other end of the
pump tubing.
7. Attach the capillary tubing to the sample inlet of the nebulizer.
Connecting the Drain Tubing Assembly
To connect the drain tubing assembly, use the following steps and refer to the next
figure.
1. Connect the sleeve (Part No. N0695415) and adapter (Part No. N0695414) to
the spray chamber drain. Thread the barbed fitting (Part No. 09920267) into
the coupler and adapter.
2. Attach the Solvent Flex (PVC) tubing (Part No. 09985735) to the barbed
fitting. Attach the other end of the Solvent Flex (PVC) tubing to the tubing
coupler (Part No. 09920186).
3. Use the other end of the tubing coupler (Part No. 09920186) to connect the
drain tubing to the peristaltic pump.
Drain tubing with 1 mm i.d. (0.04 in.) can also be used by following these steps:
Table 5-8.
Spray Chamber Drain Spare Parts
Part No.
Drain Spare Part
09920186
Barb Fitting Coupler 3/32 Tubing
09920267
Barb Fitting Male 10-32 UNF
09985708
PTFE Tubing 0.031 x 0.016
09985735
Solvent Flex Tubing, 0.11 inches ID
N0695414
Drain Adapter
N0695415
Drain Coupling
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......