180
Installation
•
Check the Interelement Corrections (IECs) if you make any changes to the
ionization or excitation characteristics of the plasma after IEC data is collected.
This would include changing the source conditions or changing the nebulizer
or torch.
•
To compensate for spectral interferences, use Background Correction,
Interelement Correction (IEC), or Multi-component Spectral Fitting (MSF).
Confirm using interference check standards.
•
To compensate for matrix effects, use internal standardization.
Hardware Settings and Options
Below is a brief list of hardware options and settings that you must set or optimize
for particular analytical requirements.
•
Make sure you use solvent resistant tubing for organic solvents. See
.
•
When the pump is not in use, release the pressure plate and release the
tubing to prevent flat spots from forming. It is recommended to flush the
tubing with DI water at the end of an analysis to preserve tube life. If
running the instrument unattended you can set the instrument to flush at the
end of an analysis using the automatic shutdown/start-up options. It is also
recommended to enable tubing saver if the instrument is set to shutdown
automatically while running unattended.
•
Make sure that the spray chamber is being drained properly to avoid
solution filling the torch.
•
For radial viewing there is one window (Part No. N0810346). Instructions
for removing the window are given in
•
PerkinElmer offers different types of nebulizers for a wide variety of
applications. For descriptions of the nebulizers and their different uses and
advantages, see
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......