110
System Description
Specifications
ICP System Specifications
Frequency:
40 MHz, free-running
Output Power Stability:
<0.1%
Power Output:
2200 watts, computer-controllable in 1 watt increments. RF
generator is located in the left side of the instrument for efficient use of laboratory
space.
RF Shielding:
Meets all FCC requirements for RF emission.
Plasma Induction Plates
Cooling Water:
System requires a flow of 1 gal/min at 310 to 550 kPa at a
temperature between 15 °C and 25 °C. A recirculating cooling system is required
(PolyScience or equivalent is recommended).
Automatic Ignition:
Plasma ignition is computer controlled and totally
automated. The plasma can be turned on at a set time, warming up the system prior
to an analysis, and can be turned off automatically after an analysis.
Safety Interlocks:
System checks water flow, shear gas flow, argon pressures,
emergency plasma off switch, torch compartment door interlocks, torch installed
and plasma stability. The status of these interlocks is constantly monitored and text
information is displayed on the computer screen. If any interlock is interrupted, the
plasma is shut down automatically.
Gas Flow Controls
Plasma Argon Flow:
Closed-loop flow control using a proportional valve and a
measured pressure across a known resistor. Computer-controlled to regulate the
flow automatically within the range of 8 to 17.0 liter/minute in 1.0 liter/minute
increments.The flow system is interlocked to prevent ignition without plasma gas
flowing.
Auxiliary Argon Flow:
Closed-loop flow control using a proportional valve and
a measured pressure across a known resistor. Computer-controlled to regulate the
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......