Exhaust Vent Requirements
•
It will remove dissipated heat that is produced by the ICP torch and RF power
supply. The venting system should provide a flow rate of at least 3398 L/min
(120 CFM). The temperatures of the exhaust gases upon exiting the instrument
are about 200 °C (392 °F) at 1500 watts of RF power.
The venting system should provide a flow rate of at least 3398 L/min (120 CFM).
The temperatures of the exhaust gases upon exiting the instrument are about 200
°C (392 °F) at 1500 watts of RF power.
The exhaust duct is connected directly to the top of the Avio 500 instrument and it
is recommended to use the 4-inch (10.16 cm) I.D. flexible ducting provided to
make the final connection to the instrument with stainless steel tubing for the rest
of the exhaust ducting. The blower capacity depends on the duct length and the
number of elbows or bends used to install the system. If an excessively long duct
system or a system with many bends is used, a stronger blower may be necessary
to provide sufficient exhaust volume. Alternatively, smooth stainless-steel tubing
may be used instead of flexible stainless-steel tubing where flexibility is not
required to reduce system friction loss or “drag.” A length of smooth stainless-steel
ducting has 20-30% less friction loss than a comparable length of flexible ducting.
When smooth stainless-steel tubing is used, elbows must be used to turn corners.
These elbows should turn at a center line radius of 150 mm with a maximum bend
angle of 45 degrees to reduce friction losses, and the number of elbows should be
minimized.
Warning: Toxic fumes
The use of ICP-OES instruments without adequate ventilation to outside air may
constitute a health hazard. For example, the combustion of halogenated
hydrocarbons produces toxic vapors. Extreme care should be taken that exhaust
gases are vented properly.
Avertissement: Des fumées toxiques
L'utilisation d'instruments d'ICP-OES sans ventilation adéquate à l'air extérieur
peut constituer un danger pour la santé. Par exemple, la combustion
d'hydrocarbures halogénés produit des vapeurs toxiques. Un soin extrême doit
être pris que les gaz d'échappement sont évacués correctement.
!
Warning
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......