Cyclonic Spray Chamber
Cleaning the Cyclonic Spray Chamber
If you have been analyzing aqueous solutions, the spray chamber should be cleaned
in an acid bath. An ultrasonic bath may be used. Start with a solution of 5% nitric
acid or aqua regia and, if deposits persist, increase the acid concentration, up to
20% if necessary. Acid solutions must be used with care. Observe all of the
manufacturer's safety recommendations. These recommendations are normally
provided in a material safety data sheet (MSDS) supplied with the chemical.
If you have been analyzing organic solutions, the spray chamber can be cleaned
using a solvent or diluted soap solution.
O-rings may be cleaned with soap and water. An ultrasonic bath may be used.
Installing the Cyclonic Spray Chamber
See the procedure on Connecting the Cyclonic Spray Chamber on page 155.
Caution
Attention
Do not use solvents that will attack quartz or glass. The cyclonic spray
chamber is not HF resistant.
Ne pas utiliser de solvants qui attaquent le quartz ou le verre. La
chambre de nébulisation cyclonique est pas HF résistant.
Table 5-2.
Replacement Parts: Spray Chambers
Part No.
Description
N0790384
Scott Spray Chamber
N0791352
Baffled Cyclonic (Standard)
N0790373
HF Resistant Cyclonic Spray Chamber
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......