105
EMO Switch
EMO Switch
The
Emergency Off
(EMO) switch is the illuminated red switch on the front of the
instrument. If it is blinking slowly the instrument is in the middle of an ignition
cycle. It blinks rapidly after the Emergency Off Switch is depressed. This is an
indication that the switch has been depressed and to remind you to push the switch
again to release it. This switch shuts off the plasma in an emergency by
disconnecting the main voltage circuitry in the RF generator. To restart, you must
release the switch by pressing the switch again. Reset the RF generator using the
software Reset button. This reset can be found by opening
Diagnostics
in the
Syngistix for ICP software, going to plasma tab and selecting the
Advanced
button.(The spectrometer stays on.)
Figure 3-16.
Location of Red Emergency Plasma Off Switch.
Item
Description
1
Emergency Off Switch
1
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......