Peristaltic Pump
Peristaltic Pump
The peristaltic pump requires routine maintenance to ensure that the system will
run properly. The pump tubing should be replaced when flat spots develop (on a
daily basis or after approximately eight hours of use). When replacing the pump
tubing, the drain tubing should be replaced at the same time since uneven wear can
cause the spray chamber to back up.
Replacement Pump Tubing
Replace the pump tubing when flat spots develop (after approximately eight hours
of use). When replacing the pump tubing, both the sample and drain tubing should
be replaced at the same time since uneven wear can cause the spray chamber to
back up.The following table gives part numbers for pump tubing for the sample and
drain.
Table 5-7.
Pump Tubing
Installing the Sample and Drain Tubing
The peristaltic pump is used to pump the sample into the nebulizer and pump the
waste out of the spray chamber. Tighten the adjustment screw (turn clockwise) up
to three turns until the bubbles are visible in the drain tubing. Then add three more
turns of the adjustment screws.
Pump Tubing Solvent Type
Inner Diameter
Part No.
Standard
Inorganic
Sample: 0.76 mm (0.030 in.)
Drain: 1.14 mm (0.045 in.)
09908587
09908585
Solvent Flex
Kerosene or xylene
Sample: 0.76 mm (0.030 in.)
Drain: 1.14 mm (0.045 in.)
00473550
09923037
Silicone
Organic solvents,
including MIBK, but not
kerosene or xylene
Sample: 0.76 mm (0.030 in.)
Drain: 1.14 mm (0.045 in.)
00473552
N0691595
Viton
Kerosene or xylene, but
not
MIBK
.
Sample: Orange-Orange
0.89 mm (0.035 in.)
Drain: Yellow-Yellow
1.40 mm (0.056 in.)
09923511
09923512
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......