Connecting the Gases and Cooling Water
An air hose with 1/4-in. Swagelok fittings at each end is supplied in the hose kit
that is shipped with the instrument. A second air hose is included with the
PerkinElmer Air Dryer Filter in the Installation kit (Part No. N0770432). The
following procedure describes how to connect the PerkinElmer air compressor,
filter and regulator.
1. Install the Air Dryer Filter as described in the instructions that are included
with the filter.
2. Locate the air hose (Part No. N0770348, uses 1/4-in. Swagelok fittings at each
end) that is shipped with the Air Dryer Filter. Connect the air hose from the air
compressor to the AIR IN fitting on the Air Dryer Filter using a 1/4-in. space
collar at both ends. Make sure the air flow is in the proper direction, as
indicated by arrows on top of the filters.
3. Locate the air hose (Part No. N0770348, uses 1/4-in. Swagelok fittings at each
end) that is shipped with the instrument in the hose kit. Connect the air hose
from the AIR OUT fitting on the Air Dryer Filter using a 1/4-in. space collar
at both ends.
Connecting the PolyScience Chiller
The PolyScience
®
WhisperCool
TM
Chiller must be used to supply cooling water to
the instrument. This chiller requires a single wall receptacle. The PolyScience
®
WhisperCool
TM
Chiller comes with a complete instruction manual. You should be
familiar with this manual before proceeding.
Note
If you are using house air, it must be clean and dry. The instrument
includes an air dryer assembly (Part No. N0775325). Refer to
Connecting
the Shear Gas Supply
later in this chapter.
Note
You are required to have your electrician install the single wall
receptacle for the PolyScience
®
WhisperCool
TM
chiller.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......