284
Maintenance
nebulizer. We have found this is one of the main causes of blockage of the
nebulizers, so please ensure that the gas line to the nebulizer is clean of any
particles.
The gas line terminates in the Avio 500 quick disconnect. It is a direct replacement
for typical concentric nebulizers.
Sample Introduction / Maximizing Stability
The Mira Mist does not have any suction, so it requires a pump to supply the
sample solution. The pump speed and the quality of the pump tubing have a large
effect on the performance of the nebulizer. Try to select a pump tubing size that
allows running the pump at a higher speed (If you use the standard black-black
tubing, you may need to use a higher pump rate: for example 1.5 mL/min). So for
low sample flow, use orange green or orange blue pump tubing. Generally the
pump tubing should be changed every day for maximum stability and lowest
%RSD.
Cleaning the Nebulizer
For the longest life and best performance, wash your nebulizer by simply running
water as a sample for 10 minutes at the end of the day before shutting down the
plasma. Any other form of washing is usually unnecessary.
DO NOT
wash your
nebulizer in acid or solvents to "prevent salt build up".
DO NOT
use surfactants as
they will destroy the surface tension at the tip and performance.
Caution
Attention
The gas orifice is at the very tip of the nebulizer. It is made of Teflon which
is
VERY SOFT
. This tip is very easily damaged and should
NEVER
be
touched with fingers, tissues, or anything else. If the tip is accidentally
touched and the nebulizer continues to operate, then it is still functional,
and its use can be safely continued.
L'orifice de gaz est à la pointe du nébuliseur. Il est fait de téflon qui est très
doux. Cette astuce est très facilement endommagé et ne doit
JAMAIS
être
touché avec les doigts, les tissus, ou toute autre chose. Si la pointe est
accidentellement touché et le nébuliseur continue à fonctionner, alors il est
encore fonctionnel, et son utilisation peut se poursuivre en toute sécurité.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......