271
GemCone Nebulizer
GemCone Nebulizer
Daily Cleaning of the Low-Flow GemCone Nebulizer
Each day after use, the low-flow GemCone nebulizer should be rinsed out by
pumping deionized water to the nebulizer for several minutes. This should be done
with both the plasma on and the nebulizer gas on. This helps prevent the formation
of salt crystals near the exit of the nebulizer gas orifice. The plasma should be on
to prevent aerosol from condensing on the purge window.
Daily Cleaning of the High Solids GemCone Nebulizer
It is important to rinse the high solids nebulizer after use, particularly when
working with solutions containing high levels of solids. You do this by pumping
deionized water for several minutes to wash out any solid material that may have
crystallized on the inside surface of the nebulizer. If a stronger cleaning solution is
required, a solution of 1% acid in deionized water may be used in place of the
deionized water. The plasma should be on to prevent aerosol from condensing on
the purge window.
To rinse the argon orifice:
1. Place the sample capillary tube in deionized water or 1% acid in deionized
water and operate the pump.
2. Momentarily shut off the nebulizer gas flow for about 30 seconds.
3. Turn the nebulizer gas back on.
This procedure can be performed more than once a day in cases where very
high levels of dissolved solids are aspirated through the nebulizer for
prolonged periods of time.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......