330
Troubleshooting
•
Organics.
If you analyzing organic solutions and cannot ignite the plasma, run
the nebulizer argon for a couple of minutes with the pump
off
to purge the spray
chamber.
•
Organic Vapors.
If these remain in the torch from an earlier analysis, they can
sometimes cause ignition problems. If you suspect this, purge the sample
introduction system with argon for several minutes.
•
Spray Chamber
. Remove the spray chamber and cap the injector. Now try to
light the plasma. If the plasma now lights and is stable the ignition problem is
in the spray chamber, nebulizer or the sample introduction system. If the
problem persists check the torch O-rings and gas connections to the torch
mount.
Plasma Stability Problems
If periodic pulsations of the torch are observed:
•
Poor sample drainage may be the problem. Make sure that waste drainage is
not backing up into the spray chamber and that the waste is draining properly.
•
Check the condition of the peristaltic pump tubing. If it is flattened, stretched,
or damaged by abrasion, replace the tubing.
•
Air leaks may be causing the pulsation. Check the O-rings on the torch
assembly and ensure that the nut on the end of the torch is firmly in place. Do
not
use a tool to tighten. Make sure the nebulizer/end cap assembly fits tightly
to the spray chamber.
If using organics, you may see an irregular pulsation:
•
Be sure that the injector you are using is recommended for the types of solvents
you are using. The alumina, 1.2mm injector, is recommended for typical
organic analyses. The alumina injector, 0.80 mm i.d., should be used with
volatile organic solvents, for example, methanol.
•
Try increasing the RF power, using less solvent, slowing down the pump
speed, or adjusting the torch height.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......