62
Preparing Your Laboratory
Additional recommendations on the venting system include:
•
Make sure the duct casing is installed using fireproof construction. Route ducts
away from sprinkler heads.
•
Locate the blower as close to the discharge outlet as possible. All joints on the
discharge side should be airtight, especially if toxic vapors are being carried.
•
Equip the outlet end of the system with a back draft damper and take the
necessary precautions to keep the exhaust outlet away from open windows or
inlet vents and to extend it above the roof of the building for proper dispersal
of the exhaust.
•
Equip the exhaust end of the system with an exhaust stack to improve the
overall efficiency of the system.
•
Make sure the length of the duct that enters into the blower is a straight length,
at least ten times the duct diameter. An elbow entrance into the blower inlet
causes a loss in efficiency.
•
Provide make-up air in the same quantity as is exhausted by the system. An
“airtight” lab will cause an efficiency loss in the exhaust system.
•
Ensure that the system is drawing properly by releasing smoke into the mouth
of the vent. A synthetic “smoke” can be generated by placing open bottles of
hydrochloric acid and ammonium hydroxide in the proximity of the vent
opening.
•
Equip the blower with a pilot light located near the instrument to indicate to the
operator when the blower is on.
Vent Positions
The venting system for the ICP torch connects directly to the torch compartment
chimney located on top of the sample compartment. 4-inch inner diameter tubing
is provided with the Avio 500 ICP-OES for making this connection. For proper
instrument venting, order PerkinElmer Venting Kit (Part No. N0790188, 110V;
N0790189, 230V).
Note
Local electrical codes do not allow PerkinElmer Service Engineers to install the
blower and vent assembly.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......